CONTEC
CMS8000
CMS 81

Features

1)12.1'' TFT color LCD, multi-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakh, Polish, Czech).
2)Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
3)All-round monitor for adult, pediatric and neonate.
4)With standard interface, oxygen graph, trend graph, big character interface and view bed, convenient to observe.
5)Finish all operations by keys and knobs.
6)Maximum 8-channel waveform synchronous display.
7)Display 7-lead ECG waveform on one screen, cascade ECG waveform display
8)Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
9)Heart rate variability (HRV) analysis function
10)NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
15)AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
Weight - 4.5 kg approx

Performance

ECG:
Multi-parameter Portable Cardiac Patient Monitor FDA
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: 5-lead, dual-channel
3-lead, single-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
RESP:
Method: R-F(RA-LL) Impedance
Respiration rate: Measurement and alarm range: 0~150 rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s

NIBP:
Method: Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult mode: 297±3 mmHg
Pediatric mode: 240±3 mmHg
Neonatal mode: 147±3 mmHg
Spo2:
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified
PR Measurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greater

TEMP:
Channel: dual-channel
Measurement and alarm range: 0 ~ 50ºC
Resolution: 0.1 ºC
Accuracy: ±0.1 ºC
EtCO2 (Optional):
Method: Sidestream or Mainstream
Measuring Range: 0~150mmHg
Resolution:0~69 mmHg, 0.1 mmHg
70~150 mmHg, 0.25 mmHg
Accuracy:0~40 mm Hg ±2 mm Hg
41~70 mm Hg ±5%
71~100 mm Hg ±8%
101~150 mm Hg ±10%
AwRR Range: 2~150 rpm
AwRR Accuracy: ±1BPM
Apnea Alarm: Yes
IBP (Optional):
Channel: dual-channel
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range: -10~300 mmHg
Resolution 1 mmHg
Accuracy: ±2% or 1mmHg, whichever is greater
Power supply: 100-240V~, 50/60Hz
Safety classification: Class I, type CF defibrillation-proof applied part